For stationary dew point monitoring of refrigeration or adsorption dryers. The touch screen graphic display enables an intuitive operation and shows the progress of the measured values. 2 alarm relays are available for monitoring of threshold values. Available either with a classic analogue output 4…20 mA or optionally with digital interfaces like Ethernet and RS 485 (Modbus protocol). As a stand-alone solution the measured data stored in the optional data logger can be read out via USB stick and evaluated by means of the software CS Soft Basic.
- USB interface
- 3.5″ graphic display with touch screen
- Integrated mains unit for supply of the sensors
- 4…20 mA output of all connected active sensors
- Pulse output (for total consumption) in case of flow sensors
- 2 alarm relays (pot.-free switch-over contacts, max. 230 V, 3 A)
- Integrated webserver
- Mathematic calculation function
- Totalizer function
- Integrated data logger
- Ethernet / RS 485 interface
- additional sensor inputs (digital or analogue) selectable
Features of the dew point measurement DS 400-Set:
- 3.5″ graphic display – easy operation with touch screen
- System ready for plug-in: Everything completely wired
- 2 alarm contacts (230 VAC, 3 A) pre- and main alarm freely adjustable
- An alarm delay can be set for each alarm relay
- 4…20 mA analogue output
- Option: Ethernet and RS 485 interface (Modbus protocol)
- Option: Webserver
Option: Integrated data logger
- Recording of the dew point progression of up to 100 million measuring values
- CS Soft Basic for evaluation in graphic and table form. Read-out of the data either via USB stick or via Ethernet
Specs
Contact Us Today To Enquire About Our Products And Services
Product Specification
Below is the product specification of the DS 400 Set – Stationary dew point measurement in compressed air systems from CS Instruments. If you require further information about the DS 400, please contact our expert customer service advisors on 01722 442 307.
Dimensions
118 x 115 x 98 mm
IP 54 (wall housing)
92 x 92 x 75 mm (panel mounting)
Inputs
2 digital inputs for FA 510 or VA 500/520
Interface
USB interface
Power supply
100…240 VAC, 50-60 Hz
Accuracy
See FA 510
Alarm outputs
2 relays, (pot.-free)
Options:
Data logger
100 million measured values start/stop time, measuring rate freely adjustable
2 additional sensor inputs
For connection of pressure sensors, temperature sensors, clamp-on ammeters, third-party sensors with 4…20 mA, 0 to 10 V, Pt 100, Pt 1000
Measuring range
-80…20 °Ctd or
-20…50 °Ctd
Accuracy
± 1 °C at 50…-20 °Ctd
± 2 °C at -20…-50 °Ctd
± 3 °C at -50…-80 °Ctd
Pressure range
-1…50 bar, special version
350 bar