Traceable Calibration

Traceable Flow Meter Calibration and Repair Services to Maintain Accuracy

To ensure optimal accuracy throughout its lifespan, it is recommended that a flow meter undergo regular calibration. Without periodic flow meter calibration, the accuracy can drift, leading to inconsistent measurements and diminished performance, which may ultimately impact your processes and product quality.

When you send us devices for calibration, we kindly ask you to either send* or enclose the following forms:

*If you choose to send the completed forms by email, please address to sales@iCenta.co.uk.

prover

iCenta Controls Calibration Laboratory 29592

Accredited in accordance with International Standard ISO/IEC 17025:2017 for the calibration of liquid flow.

metri-cert_2

UKAS Schedule of Accreditation for iCenta Controls Ltd Calibration Laboratory 29592

Piston Prover method – Meter size up to 2-inch/50mm bore.

Calibration Points

We are normally asked for either 3- 5- or 10- point calibrations but we can accommodate up to 32 points if required.

Calibration Process Medium

We can offer a range of process media for volume flow rate calibration, but we find most customers request the use of water. Other process media available include Oils and Hydrocarbons covering a viscosity range of 0.6 to 100 mm²/s (cSt). For mass flow rate we only offer water as the process media.

iCenta Controls Traceable Calibration 

This is not part of our UKAS accreditation, but measurements are traceable to national and international standards.

Piston Prover method – Meter size up to 3-inch/ 75 mm bore.

Calibration Points

We are normally asked for either 3- 5- or 10- point calibrations but we can accommodate up to 32 points if required.

Calibration Process Medium

We can offer a range of process media for volume flow rate calibration, but we find most customers request the use of water. Other process media available include Oils and Hydrocarbons covering a viscosity range of 0.6 to 100 mm²/s (cSt).

Measured Quantity Instrument or Gauge
Range
 Measurement Uncertainty
Remarks
LIQUID FLOW
Piston prover method Calibration fluid:
Flow rate – volume
0.01 l/min to 1100 l/min
0.1%
Water, Oils and Hydrocarbons
Flow rate – mass
0.01 kg/min to 1100 kg/min
0.15%
Water, Oils and Hydrocarbons

Master Meter method – Meter size 2-inch (25mm) to 8-inch (200mm) bore.

Calibration Points – Normally either 3- 5- or 10- point calibrations.

Calibration Process Medium – Water.

Measured Quantity Instrument or Gauge
Range
Measurement Uncertainty
Remarks
LIQUID FLOW
Master Meter method Calibration fluid:
Flow rate – volume
12 m3/h to 220 m3/h
0.5%
Water

​Some of the many types of flow technology we calibrate, and service are as follows;

Electromagnetic (Mag), Turbine, Oval Gear, Positive Displacement, Coriolis Mass, Vortex, Ultrasonic, Thermal, Paddlewheel (Pelton Wheel), Variable Area (Rotameter)

Need Advice on the best way to calibrate your flow meter?